Laminar Flow Clean Bench
VLF 797/HLF 797
SPECIFICATIONS
| Filters |
| Aluminum framed High Efficiency Particulate Air (HEPA) type,
with a minimum efficiency of 99.99% at 0.3 micron, aerosol
challenged (zero probed). |
| Prefilters |
| Washable type, polyurethane foam - 30% ASHRAE. |
| Airflow |
| 90 feet per minute ±10 FPM (0.45 meter per second)
average velocity measured 6 in. (152.4 mm) from the diffuser
screen. Uniformity ±20% of average or better. |
| Motor/Blower
Assembly |
Direct drive, continuous duty 1/5 Hp, 300W, with sealed-forlife
bearings and inherent overload protection. Motor/blower
assembly is electronically balanced (dynamically) and
mounted on vibration dampening rubber mounts.
Motor/blower assembly is designed to provide rated airflow
through a 50% increase in initial static pressure. |
| Speed
Controller |
| Solid-state speed control with RFI suppression for adjusting
airflow velocity utilizing a, easy to clean, membrane touchpad controller. |
| Electrical |
Standard 15 Amp, 115V, single phase, 60 Hz, rubber covered
cord with 3-prong grounded plug.
Single electrical outlet flush mounted on front blower
housing.
Optional fluorescent lighting fixture provides 150 foot-candles
(1615 lux) at the work surface.
|
| Construction |
Aluminum cabinet with front panel finished in corrosion
resistant polyurethane enamel. |
| Warranty |
| Limited
one year. |
| Miscellaneous |
Power requirements - 115V, 60hz - 3.0 FLA
Heat Gain - 1000 BTUH
Airflow - 650 CFM @ 90 FPM |