Hospi-Gard®
AirCeil™
SPECIFICATIONS
| Motor/Blower
Assembly |
| Two
direct drive, continuous duty 128 Watt motors with
sealed-for-life bearings and inherent overload protection
located on the "clean side" of the filter. Fan, dual inlet,
centrifugal blower. The motor/blower assembly is mounted on
vibration dampening rubber mounts and designed to provide
up to 650 CFM of air at a filter pressure drop from 0.35 to
1.00 inch w.g. |
|
Electrical |
Standard
115 V, single phase, 50/60 Hz with 13/16-in. conduit knock-out.
Standard
220 V, 50 Hz with 20.6mm conduit knock-out. |
| Construction |
| Welded
aluminum filter plenum and blower housing with
exterior room-side surface finished in a corrosion resistant
polyurethane enamel. |
| Warranty |
| Limited
one year. |
| Hospital
Applications |
- Negative
Pressure Rooms
-
Waiting Rooms
- Aerosol
Pentamidine Treatment
- Bronchoscopy
Rooms
- Emergency
Rooms
- Sputum
Induction
- Intensive
Care Units
- Renal
Dialysis Rooms
|
| Other
Applications |
- Clinics
-
Physician Offices
-
Addiction Recovery Centers
-
Nursing Centers
-
Homeless Shelters
- Correctional
Facilities
|
| Options |
- AirCeil™
with UV-C germicidal light, Part No. 11122-002
-
Infared Remote Controller (one required), Part No.63760-002
-
Field Installed UV-C Light Part No. 24375-001
-
Hospital Grade Power Cord, Part No. 10660
-
Carbon Prefilter, Part No. 24376-001
-
1/4 in.-20 Eyebolts (set of 4), Part No. 62529-001
-
Ceiling Air Diffuser, Part No. 63857
-
12 in. (305 mm) Flexible Duct, Part No. 63856
-
Room Pressure Monitor, Part No. 63337
-
Annunciator, Part No. 63338
|
| Part
Number |
11122-001 |
 |
| Airflow,
CFM (m3/hr) |
Variable
400 to 650 (680-1105) |
| Maximum
BTU / Hour |
875 |
Power
Requirements
Amps @ 115 V / 60 Hz
(230 V/50Hz) |
400
CFM: 0.72 Amps |
| 650
CFM: 1.20 Amps |
| Full
Load Amps: 2.25 (1.0) |
| UL
Listed |
Yes |
Dimensions,
inches
(millimeters) |
A |
47.62
(1210) |
| B |
19.5
(496) |
| C |
23.62
(600) |
| D |
3.00
(77) |
Unit
Weight, lbs.
(kilograms) |
85
(38.6) |
For
additional information,
please contact your local Representative or
Distributor.
SPECIFICATIONS
ARE SUBJECT TO CHANGE WITHOUT NOTICE AND WITHOUT INCURRING LIABILITY FOR